0-15L Minilator oxygen gas diverter valve
The LSP Minlator is a compact lightweight
and simple-to-operate constant flow device which
provides oxygen at flow rates up to 15 lpm for
1 to 5 patients simultaneously.The unit is
comprised of an anodized aluminum manifold
with check valves and threaded barbed fittings
with metered orifices.
The LSP Minlator is attached via an oxygen supply
hose directly to an oxygen regulator capable of
delivering 40 to 90 psig.Several Minlators may
be connected in tandem for situations such as
disasters where more than 5 oxygen outlets would
be required.The Minlator also provides a 50 psig
outlet which can be used to operate a demand valve
resuscitator or a suction device.
Orifice Pressure vs.Flow Rate | |||||
Oxygen PSI | 2 Ipm | 4 Ipm | 6 Ipm | 10 Ipm | 15 Ipm |
40 psi | 1.4 lpm | 2.6 lpm | 3.8 lpm | 6.6 lpm | 9.7 lpm |
50 psi | 1.6 lpm | 3.01pm | 4.5 lpm | 7.8 lpm | 11.7 lpm |
60 psi | 1.9 lpm | 3.5 lpm | 5.2 lpm | 8.9 lpm | 11.7 lpm |
70 ps | 2.1 lpm | 3.5 lpm | 5.2 lpm | 8.9 lpm | 13.8 lpm |
80 psi | 2.4 lpm | 4.0 lpm | 5.9 lpm | 10.1 lpm | 15.7 lpm |
90 psi | 2.6 lpm | 5.0 lpm | 7.2 lpm | 12.6 lpm | 19.3 lpm |
Flow Tolerance Accuracy | .4 lpm | .8 lpm | 1.2 lpm | 2.0 lpm | 3.0 lpm |
L419-010 |
Minlator:Aluminum manifold with 5 oxygen DISS check valves, five screw-on 4 lpm orifices,zippered nylon carry case |
L419-020 | Minlator:same as L419-010 but with five 6 lpm orifices |
L419-030 | Minlator:same as L419-010 but with five 10 lpm orifices |
L419-050 | Minlator:same as L419-010 but with five 15 lpm orifices
|
SPECIFICATIONS
Storage Temperature:-40°F to 160°F
Operating Temperature:-30°F to 125°F
Supply Pressure:40 to 90 psig
Flow Rates:2,4,6,10 or is lpm
Weight:13.5 ounces
Dimensions without barbs:1″x3″x 6″
Dimensions with barbs attached:1″x 6″x 7.5″
Manifold:Anodized aluminum
Fittings:Plated brass Barbs:H.I.polystyrene
Orifices:Stainless steel